The R-22A Sensor with 3 Pin molex
We provide a 12 Month Warrenty for our oxygen sensors.
This replacement sensor can be used e.g. with the following analyzers:
Productspezification:
Electr. Connection | 3-Pin Molex |
| Output range: | 0 - 100 Vol. % |
| Expected life: | 36 months at 20.9% oxygen, 25° C, 50% RH |
| Output range: | 7 - 13 mV @ 20,9% O2 |
| Response time: | actual response of sensor is < 6 seconds 0 to 90% O2 |
| Linearity: | < 2 Vol.% +/- 0.1% < 25 Vol.% +/- 0.5% < 100 Vol.% +/- 1% |
| Operating temperature range: | 0 - 40°C |
| Luftdruckbereich: | 600 - 1750 hPa |
| Linearity: | 10% pro Jahr |
| Signalbeeinflussung durch N2: | Ua/200 @ 100% N2 in < 40 sec |
| Humidity interference: | -0.03% rel. O2 Anzeige pro % r.F. |
| Temperature compensation: | NTC |
| Cross interference: | < 0.1 % O2 Beeinflussung durch: 15.0 % CO2 balance N2 10.0 % CO balance N2 3,000 ppm NO balance N2 3,000 ppm C3H8 balance N2 500 ppm H2S balance N2 500 ppm SO2 balance N2 1,000 ppm Benzol balance N2 |
Storage:
| Operating temperature range | 0 - 40°C |
| MAximum Operating temperature range: | -20 - 50°C |
| Atmospheric pressure: | 600 - 1750 hPa |
| Humidity:: | 0-99% |
| shelf life: | weniger als 6 Monate |
| weight: | ca. 32 g |
| used materials: | PA 12, PPS, PTFE, Edelstahl |
| warrenty: | 12 Monate |
| colour | Weißes Gehäuse |
Alle Angaben basieren auf 25°C, 50% rel. H. und 1013 mBar Luftdruck